AFC123 Personal Sampler. Page 21
Handbook No. HB3182-02 Issue2/JULY 1994
Possible cause:-
Service Hint.
Grub screw securing eccentric to motor Re tighten
shaft loose.
Symptom:- Pump runs but no air drawn.
Possible cause:-
Service Hint.
Diaphragm rubber split. Replace.
Pulsation damper membrane split. Renew pulsation
damper.
Flow potentiometer on bottom setting. Re-adjust.
Tube between pulsation damper and Replace.
valve chest split or disconnected.